Raw EMG App

This application collects and processes raw EMG and IMU data during activities. The app interfaces with a sensor device to read EMG signals, apply filtering, and calculate outputs for visualization and storage.

Nodes Required: 2

  • Sensing (2):

    • EMG Sensor (attached to skin on target muscle )

    • IMU Sensor (attached to the body or equipment)

  • Feedback (0)

Algorithm Information

The raw EMG data from the sensor undergoes a series of filtering steps:

1. Bandpass Filtering: Removes unwanted low- and high-frequency noise based on specified frequency bounds.

2. Notch Filtering: Removes power line interference (50 or 60 Hz).

3. RMS Envelope Calculation: Calculates the root-mean-square (RMS) envelope using a moving average approach.

Processed Data Variables

- raw_data: Original unprocessed EMG data.

- bandpassed_data: EMG after bandpass filtering.

- notched_data: EMG after notch filtering.

- envelope_data: Final processed RMS envelope of EMG.

User Configurable Settings

Band Filter Settings

  • Low Cut Frequency: Default 10 Hz

  • High Cut Frequency: Default 100 Hz

  • Notch Filter Frequency: Selectable (50 or 60 Hz)

Save Options

  • Output File Format: CSV, H5, XLSX

These settings directly influence data processing and output storage.

Description of Data in Downloaded File

Saved and Calculated Fields

  • Time (s): Time elapsed since the start of collection.

  • Raw_EMG (mV): Unprocessed EMG sensor data.

  • Bandpass_Filter (mV): EMG data after bandpass filter.

  • Notch_Filter (mV): EMG data after notch filter.

  • RMS_Envelope (mV): RMS envelope of the filtered EMG.

Data is recorded at 500Hz and streamed at 100Hz for visualization.

IMU Data

  • AccelX/Y/Z (m/s²): Raw acceleration data from the IMU sensor.

  • GyroX/Y/Z (°/s): Raw gyroscope data from the IMU sensor.

  • MagX/Y/Z (μT): Raw magnetometer data from the IMU sensor.

  • Quat1/2/3/4: Quaternion data for sensor orientation.

  • Sampletime: Timestamp for each sensor data point.

  • Package: The package number associated with the sensor data.